8. After the deuterium lamp cools, remove the lamp and discard it.
CAUTION!
Do not touch the glass of the new lamp. Fingerprints on the glass can
shorten the life of the lamp.
9. Being careful not to touch the glass bulb with bare hands or fingers, line up the
alignment notch in the base of the lamp with the guide on the lamp block to position the
new lamp correctly.
10. Use the flat head screwdriver to tighten the flat head screws to secure the deuterium
lamp.
11. Reconnect the lamp power cable. Make sure that the locking clip on the connector is
latched.
Note:
The lamp power cable connector is keyed to prevent incorrect connection.
12. Replace the lamp housing cover and use the hex key to tighten the two hex screws that
secure it in place.
13. Reconnect the power cable to the instrument.
14. Power on the instrument.
Packing the Instrument
To minimize the possibility of damage during storage or shipment, the instrument should be
repacked only in the original packaging materials. To order replacement packing materials,
contact your Molecular Devices representative.
Correctly repacking the instrument includes following applicable decontamination
procedures and installing the transport lock on the plate drawer.
CAUTION!
When transporting the instrument, warranty claims are void if improper
packing results in damage to the instrument.
Store the instrument in a dry, dust-free, environmentally controlled area.
Illustration
Part Number
Description
YW 000 006
2.0 mm hex key
Required Tool
CAUTION!
Do not touch or loosen screws or parts other than those specifically
designated in the instructions. Doing so could cause misalignment and possibly void
the warranty.