![MOGlabs FSC Manual Download Page 39](http://html.mh-extra.com/html/moglabs/fsc/fsc_manual_1823377039.webp)
B. Troubleshooting
B.1
Laser frequency not scanning
A
MOGL
abs
DLC
with external piezo control signal requires that
the external signal must cross 1.25 V. If you are sure your external
control signal crosses 1.25 V confirm the following:
•
DLC
span is fully clockwise.
•
FREQUENCY
on the
DLC
is zero (using the
LCD
display to set
Frequency
).
•
DIP9
(
External sweep
) of the
DLC
is on.
•
DIP13
and
DIP14
of the
DLC
are off.
• The lock toggle switch on the
DLC
is set to
SCAN
.
•
SLOW OUT
of the
FSC
is connected to the
SWEEP / PZT MOD
input of the
DLC
.
•
SWEEP
on the
FSC
is
INT
.
•
FSC
span is fully clockwise.
• Connect the
FSC MONITOR 1
to an oscilloscope, set the
MONI-
TOR 1
knob to
RAMP
and adjust FREQ OFFSET until the ramp
is centred about 1.25 V.
If the above checks have not solved your problem, disconnect the
FSC
from the
DLC
and ensure that the laser scans when controlled
with the
DLC
. Contact
MOGL
abs for assistance if not successful.