8
Chapter 2. Connections and controls
STATUS
Multi-colour indicator displaying status of the lock.
Green
Power on, lock disabled.
Orange
Lock engaged but error signal out of range, indicating the lock
has failed.
Blue
Lock engaged and error signal is within limits.
2.1.5 Signal monitoring
Two rotary encoders select which of the specified signals is routed to
the rear-panel
MONITOR 1
and
MONITOR 2
outputs. The
TRIG
output
is a
TTL
compatible output that switches from low to high at the
centre of the sweep. The table below defines the signals.
CHA
Channel A input
CHB
Channel B input
FAST ERR
Error signal used by the fast servo
SLOW ERR
Error signal used by the slow servo
RAMP
Ramp as applied to
SLOW OUT
BIAS
Ramp as applied to
FAST OUT
when
DIP3
enabled
FAST
FAST OUT
control signal
SLOW
SLOW OUT
control signal