![MOGlabs FSC Manual Download Page 15](http://html.mh-extra.com/html/moglabs/fsc/fsc_manual_1823377015.webp)
2.3 Internal DIP switches
11
2.3
Internal DIP switches
There are several internal
DIP
switches that provide additional op-
tions, all set to
OFF
by default.
WARNING
There is potential for exposure to high voltages inside the
FSC
, es-
pecially around the power supply.
OFF
ON
1
Fast gain
Front-panel knob
External signal
2
Slow feedback Single integrator
Double integrator
3
Bias
Ramp to slow only
Ramp to fast and slow
4
External
MOD
Disabled
Enabled
5
Offset
Normal
Fixed at midpoint
6
Sweep
Positive
Negative
7
Fast coupling
DC
AC
8
Fast offset
0
+2.5 V
DIP 1
If
ON
, fast servo gain is determined by the potential applied to the
rear-panel
GAIN IN
connector instead of the front-panel
FAST GAIN
knob.
DIP 2
Slow servo is a single (
OFF
) or double (
ON
) integrator. Should be
OFF
if using “nested” slow and fast servo operation mode.
DIP 3
If
ON
, generate a bias current in proportion to the slow servo output
to prevent mode-hops. Only enable if not already provided by the
laser controller. Should be
OFF
when the
FSC
is used in combination
with a
MOGL
abs
DLC
.
DIP 4
If
ON
, enables external modulation through the
MOD IN
connector
on the rear panel. The modulation is added directly to
FAST OUT
.
When enabled but not in use, the
MOD IN
input must be terminated
to prevent undesired behaviour.
DIP 5
If
ON
, disables the front-panel offset knob and fixes the offset to
the mid-point. Useful in external sweep mode, to avoid accidentally