25. Technical Data
Measuring ranges
with sensor 1mV/ms
-2
Human vibration:
Acceleration:
Velocity:
Displacement:
with sensor 10mV/ms
-2
Human vibration:
Acceleration:
Velocity:
Displacement:
0.10-12.00 / 1.0-120.0
*
/ 10-1200 / 100-6000 m/s²
peak
0.10-12.00 / 1.0-120.0
*
/ 10-1200 / 100-6000 m/s²
peak
0.001-0.120 / 0.010-1.200 / 0.10-12.00 / 1.00-60.00 m/s
peak
0.001-0.120 / 0.010-1.200 / 0.10-12.00 / 1.00-60.00 mm
peak
0.010-1.200 / 0.10-12.00
*
/ 1.0-120.0 / 10-600 m/s²
peak
0.010-1.200 / 0.10-12.00
*
/ 1.0-120.0 / 10-600 m/s²
peak
0.10-12.00 / 1.0-120.0 / 10-1200 / 100-6000 mm/s
peak
0.10-12.00 / 1.0-120.0 / 10-1200 / 100-6000 µm
peak
Full-scale values valid only for frequencies where filter or
integrator attenuation is 0 dB.
* Reference range for calibration (chapter 15.3)
Accuracy:
± 3 % and ± 2 digits at reference conditions
Linearity error:
< 5% of reading in all measuring ranges
Display modes:
Running RMS (1 s)
Maximum running RMS (MTVV)
Interval RMS (up to 10 hours)
Estimated vibration dose value (eVDV)
Vibration total value (A
hv
)
Peak value (1 s)
Maximum peak value
Crest factor
Weighting filters:
W
b
, W
c
, W
d
, W
e
, W
g
, W
h
, W
j
, W
k
, W
m
Band filters (3 dB limits):
6 – 200 Hz (displacement)
0.4 – 100 Hz (acceleration and velocity)
2 – 300 Hz (acceleration and velocity)
0.4 – 1250 Hz (acceleration and velocity)
10 – 1250 Hz (acceleration and velocity)
0.4 – 10 000 Hz (acceleration)
1000 – 10 000 Hz (acceleration)
The filters have 2
nd
order Butterworth characteristics.
Display:
Graphical LCD with 32 x 120 dots and LED backlight
3 vibration values with unit and mode
Sensor inputs:
3 IEPE inputs, socket type
Binder
711, female, 4 pins
IEPE supply:
3 constant current sources, 2 mA, compliance voltage 20V
automatic sensor detection
maximum input voltage: 6 V (peak-peak)
Transducer sensitivity:
Nominal, adjustable 0.8 – 1.2 mV/ms
-2
Nominal 10 mV/ms
-2
, adjustable 8 – 12 mV/ms
-2
Recommended Accelerometers: KS943B.10 for Hand-Arm measurement (1 mV/ms
-2
)
KB103SV-100 for whole-body measurement (10 mV/ms
-2
)
Cross-talk between X / Y / Z:
< 0.5 % at ± 3 V input voltage, filter 0.4 – 10 000 Hz
Signal output:
3 AC sensor signals, buffered and unfiltered,
peak output voltage ± 2 V, output impedance < 100 Ω
Socket type
Binder
711, male, 4 pins
User-defined setups:
10 individually selectable parameter sets for vibration
quantity, filter and display mode
50