971B UniMag™
24
Operation manual
Status Query Commands
Device Type - DT
Specifies transducer device type name:
Query:
@253DT?;FF
Query reply:
@253ACKUNIMAG;FF
Firmware Version - FV
Specifies transducer firmware version: Query:
@253FV?;FF
Query reply:
@253ACK1.12;FF
Hardware Version - HV
Specifies transducer hardware version:
Query:
@253HV?;FF
Query reply:
@253ACKA;FF
Manufacturer - MF
Specifies transducer manufacturer:
Query:
@253MF?;FF
Query reply:
@253ACKMKS;FF
Model - MD
Specifies transducer model number:
Query:
@253MD?;FF
Query reply:
@253ACK971B;FF
Part Number - PN
Specifies transducer part number:
Query:
@253PN?;FF
Query reply:
@253ACK971B-11030;FF
Serial Number - SN
Specifies transducer serial number:
Query:
@253SN?;FF
Query reply:
@253ACK0825123456;FF
Time ON - TIM
The TIM command returns the number of hours the transducer has been on:
Query:
@253TIM?;FF
Query reply:
@253ACK123;FF
Cold Cathode Time ON – TIM2
The TIM2 command returns the number of hours the cold cathode high voltage has been on:
Query:
@253TIM2?;FF
Query reply:
@253ACK123;FF
Cold Cathode pressure dose – TIM3
The TIM3 command returns the cold cathode pressure dose. (Refer to page 17)
Query:
@253TIM3?;FF
Query reply:
@253ACK1.23E-4;FF
Transducer Status – T
The T command returns the Transducer sensor status as O for OK, C for Cold Cathode failure or G for Cold
Cathode ON.
Query:
@253T?;FF
Query reply:
@253ACKO;FF