4. Analyser Installation
This section deals with getting the equipment you have just bought out of its box
and installing it on your system. If you have any questions or experience any
difficulties, contact your local MKS Spectra representative who will be able to help
you.
4.1 Unpacking
When you receive the e-Vision
carefully check each item before removing the foam
packaging and plastic wrapping, to ensure that no physical damage has occurred
during shipment. Also, make sure all items have been received correctly by
checking each item against the enclosed packing slip.
If there has been obvious damage during shipment, or if there are items listed on
the packing slip as shipped which are not in the box, immediately contact your local
MKS Spectra sales/service representative.
4.2 Inspecting the analyser
CAUTION
The analyser is both fragile and very easily contaminated by the
slightest touch from your fingers or undesirable surfaces.
Once you have inspected the analyser, return it to the protective
envelope until you are ready for installation.
The analyser is supplied vacuum sealed inside a vac-formed plastic envelope for
protection. Carefully unfold the envelope to allow access to the analyser.
Do not discard this packaging.
Hold the analyser ONLY at the vacuum flange.
Hold the analyser at the flange, with the quadrupole and ioniser structure vertically
up. Carefully inspect all the observable insulators for damage.
Look at each lead from the flange to its termination point to ensure that it does not
touch any other element of the analyser.