Appendix A: Product Specifications
50
RoHS (Restriction of Hazardous
Substances) Compliance
Compliant with Directive 2002/95/EC
CE Compliance
Electromagnetic Compatibility
Compliant with EMC Directive 2004/108/EC
Temperature Coefficients
Zero (≥ 10 Torr ranges)
Zero (< 10 Torr)
Span
150
C Units
0.004% F.S./°C
0.008% F.S./°C
0.02% of Reading/°C
200
C Units
0.008% F.S./°C
0.016% F.S./°C
0.02% of Reading/°C
Time Response
<50 milliseconds
Environmental Specifications
Ambient Operating Temperature Range
150
C Units
200
C Units
10° to 50° C* (50° to 122° F)
* Minimum 50 ft/min air flow is required if the ambient
temperature is between 45
to 50
C (113
to 122
F)
10° to 50° C** (50° to 122° F)
** Minimum 150 ft/min air flow is required if the
ambient temperature is between 40
to 50
C (104
to
122
F)
Storage Humidity Range
25 to 95% Relative Humidity, non-condensing
Storage Temperature Range
-20 to +80
C (-4 to 176° F)
Due to continuing research and development activities, these product specifications are subject to change without notice.
4
Includes non-linearity, hysteresis, and non-repeatability.
Summary of Contents for Baratron 631D
Page 3: ...1039281 001 Rev B 3 13 MKS Baratron Type 631D High Temperature Pressure Transducer...
Page 28: ...Chapter One General Information 20 This page intentionally left blank...
Page 36: ...Chapter Two Installation 28 This page intentionally left blank...
Page 44: ...Chapter Three Overview 36 This page intentionally left blank...
Page 50: ...Chapter Four Operation 42 This page intentionally left blank...
Page 54: ...Chapter Five Optional Solid State Relay System 46 This page intentionally left blank...
Page 61: ...Appendix B Model Code Explanation 53...
Page 62: ...Appendix B Model Code Explanation 54 This page intentionally left blank...