Chapter Three: Overview
33
Interface Connector
The Interface connector (J1) is a 15-pin male type D-subminiature connector. Refer to Figure 2:
Top View of the Type 631D Transducer, page 30, for the location of the Interface connector for
the 631D without the optional relay system. Refer to Figure 3: Top View of the Type 631D
Transducer With Optional Relay System, page 31, for the location of the Interface connector for
the 631D without the optional relay system. It contains the pins to connect the 631D unit to a
power supply or power supply/readout. Table 6 specifies the pinout of the Interface connector.
Note
The state of the relay contacts at pins 13, 14 are for an energized heater
alarm relay (normal operation) – refer to Heater Status LEDs and Heater
Alarm Relay, page 40 for a description of relay functionality. To use a
631D transducer with any non-MKS power supplies or power
supply/readouts, consult the manufacturers’ specifications for electrical
and power considerations, and Appendix A: Product Specifications
page
49
,
for electrical requirements of the 631D transducer.
Pinouts of the Interface Connector (J1)
Pin Number
Assignment Without Relay System
Assignment With Relay System
1
No Connection
Trip Point A Relay N/O
2
Pressure Output
Pressure Output
3
No Connection
Trip Point A Relay Common
4
No Connection
Trip Point A Relay N/C
5
15 V Return
15 V Return
6
-15 VDC
-15 VDC
7
+15 VDC
+15 VDC
8
No Connection
Trip Point B Relay N/O
9
No Connection
Trip Point B Relay Common
10
No Connection
Trip Point B Relay N/C
11
Heater Alarm Relay Common
Heater Alarm Relay Common
12
Pressure Output Return
Pressure Output Return
13
Heater Alarm Relay Open [when
energized]
Heater Alarm Relay Open [when
energized]
14
Heater Alarm Relay Closed [when
energized]
Heater Alarm Relay Closed [when
energized]
15
Chassis Ground
Chassis Ground
Summary of Contents for Baratron 631D
Page 3: ...1039281 001 Rev B 3 13 MKS Baratron Type 631D High Temperature Pressure Transducer...
Page 28: ...Chapter One General Information 20 This page intentionally left blank...
Page 36: ...Chapter Two Installation 28 This page intentionally left blank...
Page 44: ...Chapter Three Overview 36 This page intentionally left blank...
Page 50: ...Chapter Four Operation 42 This page intentionally left blank...
Page 54: ...Chapter Five Optional Solid State Relay System 46 This page intentionally left blank...
Page 61: ...Appendix B Model Code Explanation 53...
Page 62: ...Appendix B Model Code Explanation 54 This page intentionally left blank...