Chapter Two: Installation
23
Cable to Connect to the Power Supply
The 631D transducer is powered by a single
15 VDC power supply for both the electronics and
heater. The cable connection to the 631D transducer via the Interface connector is shown in
Figure 2: Top View of the Type 631D Transducer, page 30. Refer to Table 6: Pinout of the
Interface Connector (J1), page 34, for the complete list of pin assignments for the Interface
connector.
Product Location and Requirements
Voltage:
15 VDC (±5%)
Current: 1.0 Amp, maximum for the heater and electronics
Overpressure limits: 45 psia or 120% of sensor Full Scale, whichever is greater
Mount the Type 631D transducer components correctly: refer to Figure 1: Recommended
Orientation of the Type 631D Transducer, page 26, for the recommended orientation of the
sensor
Maintain a solid system ground for proper operation and safety to personnel
Isolate the unit from vibration
Keep the unit free from vibration. The diaphragm may be susceptible to resonance when
there is not enough gas present to dampen the effects of vibration. The 1 Torr Full Scale
units are the most sensitive. Any vibration that
does
exist should be isolated from the unit
with vibration isolation mounting and a bellows coupling.
Warm up Time: 2 hours
Summary of Contents for Baratron 631D
Page 3: ...1039281 001 Rev B 3 13 MKS Baratron Type 631D High Temperature Pressure Transducer...
Page 28: ...Chapter One General Information 20 This page intentionally left blank...
Page 36: ...Chapter Two Installation 28 This page intentionally left blank...
Page 44: ...Chapter Three Overview 36 This page intentionally left blank...
Page 50: ...Chapter Four Operation 42 This page intentionally left blank...
Page 54: ...Chapter Five Optional Solid State Relay System 46 This page intentionally left blank...
Page 61: ...Appendix B Model Code Explanation 53...
Page 62: ...Appendix B Model Code Explanation 54 This page intentionally left blank...