4
Installation
24
Series 943 Cold Cathode Vacuum Sensor
Instruction Manual - #100009887
4.1.8 Relay Inductive Loads and Arc Suppression
If the set point relays are used to switch inductive loads, e.g., solenoids, relays, transformers, etc.,
the arcing of the relay contacts may interfere with Controller operation or reduce relay contact life.
Therefore an arc suppression network (see Figure 4-4) is recommended. The values of the
capacitance
C
and the resistance
R
are calculated by the equations:
where,
a
= 1 + (
50
/E)
C
is in microfarads
R
is in ohms
I
is DC or AC
peak
load current in amperes
E
is DC or AC
peak
source voltage in volts
Note that
C
min
= 0.001 mF and
R
min
= 0.5 W.
Figure 4-4
:
Relay Arc Suppression Network
Table 4-3: Remote High Voltage Enable/Disable
If
Pin 9 is:
High voltage is:
Disconnected
or
Connected to a “logic Hi” of +5V
ON (sensor enabled)
Connected to pin 10 through a
switch or relay contact
or
connected to a “logic Low” of 0V
OFF (sensor disabled)