Page 32
SECTION III MAINTENANCE
3-1. Routine Inspections
㺃
Performed Daily
by Operator
The following inspections should be done before
and after each use of the table.
a.
Inspect all table pads for damage.
b.
Inspect all table top sections for damage.
c.
Inspect the table top assembly, all top sections
and the base for stability.
d.
Inspect the power cord and plug for any signs of
burns or damage.
e.
Test the operation of the main power switch.
f.
Test all functions of the pendant control for
proper table movement.
g.
Inspect the table base surface and the floor for
any signs of oil leaking.
h.
Perform steps of cleaning procedure.
i.
Recharge batteries /check battery charge
CAUTION
Do not use the table if any of the in-
spection points fail.
Performed weekly
a.
Check each function for movement to ensure
quiet and smooth operation.
b.
Check the overall condition of the hand control.
Performed monthly
a.
Inspect casters and hydraulic floor lock assembly,
clean as necessary.
b.
Inspect the emergency back-up switches for op-
eration.
c.
Inspect hand control cord condition.
d.
Inspect table pad condition.
3-2. Preventive Maintenance
The following preventive maintenance checks and
services are recommended to ensure the servicea-
bility and proper operation of your MIZUHO Surgical
Table, and should only be performed by qualified
MIZUHO trained personnel.
During normal cleaning, a general visual examina-
tion should be made checking for leaks, loose bolts
or parts, and cracked, chipped, or missing paint. Any
necessary repairs should be made.
Annually or as required based on usage, the fol-
lowing checks and services should be performed:
a.
Check all hydraulic fittings, mini-valves and slave
cylinders for proper operation and any signs of
leaks.
b.
Check the hydraulic speed controls and adjust if
necessary.
c.
Pressure check (with a gauge) the pressure relief
valve.
d.
Check all mechanical adjustments and adjust as
necessary.
e.
Check hydraulic fluid level.
f.
Lubricate the slider assembly.
g.
Check function of back and foot leg release lev-
ers, lubricate as necessary.
h.
Check the recharging of batteries:
㺃
Whenever table is placed into use.
㺃
Whenever the last bar on power panel is dis-
played.
㺃
Whenever the table is unused for an extended
period of time.
i.
Check table top level function.
j.
Inspect emergency back-up controls.