Page 10
SPECIAL USER ATTENTION
WARNING
Always inspect product prior to use to
ensure safe and correct operation. Any
product deemed to be malfunctioning
should be removed from service and
labeled inoperable. Refer all service to
a qualified MIZUHO service repre-
sentative.
NOTE
If any buttons on the control unit are
pressed continuously for longer than 10
minutes, the thermal protector of the
pump motor will actuate and the opera-
tion of the table will be stopped. The
thermal protector will reset in approxi-
mately 30 minutes.
NOTE
The operating table a built-in
self-diagnosis function to check if the
operating table and control unit work
properly.
1. Connect the power cord.
2. Press the BATTERY ON/OFF button
on the pendant control.
3. Press one of the function buttons on
the pendant control.
If two green battery indicators are lit,
the pendant control works properly.
NOTE
Activating any function button will acti-
vate the brake system. Using the
TABLE UP function to set the brakes
provides a visual assurance that the
brakes are locked without altering the
table position, except when emergency
brake is released.
WARNING
Prior to operating the table, observe all
table caution labels and review the
SPECIAL USER ATTENTION section in
the front of this manual.
WARNING
Possible explosion hazard exists if table
is used in the presence of
FLAMMABLE ANESTHETICS.
NOTE
An equalization terminal is located un-
der the main power panel. This is pro-
vided as an alternate pathway to re-
duce the risk of static shock hazards.
Always follow recommended grounding
procedures to ensure patient and staff
safety.
WARNING
To avoid the risk of electric shock, this
equipment must only be connected to a
supply main with Protective earth.
Power Cord has to be used the one
which has a logotype of MIZUHO.
WARNING
Please don't position the surgical table
in such way that the operator has diffi-
culty to disconnect the Equipment by
plug or appliance outlet.
NOTE
The table will operate correctly on bat-
tery power with the power cord con-
nected to a wall outlet or disconnected.
NOTE
Battery Operation must be turned OFF
at the pendant control. It cannot be
turned OFF using the main power
switch.
NOTE
Turning the Main Power Switch ON will
change the table operation to 100 - 240
VAC power.