5 - 6
No. 99MBC034B
5.2.7 Time required for RS-232C data output
The maximum output time (ms) as a result of executing the all-
data output command (GA00CRLF) is calculated as follows:
= (number of connected counters x 10) + (number of available
channels x 17) + 6
[= (number of connected counters x 10) + (number of available
channels x 8.5) + 3]
Transmission speed: 9600bps [19200bps]
(Example)
One EV-16P unit (total 6 gage channels): Max. 118ms [64ms]
Ten EV-16P units (total 60 gage channels): Max. 1126ms [610ms]
NOTE
•
"Number of connected counters x 20" is used for EF
Counters.
•
The above figures do not include personal computer
processing time.
•
The data size required if total 60 gage channels are
used, is 900 bytes (15 bytes per gage channel). For
more information refer to the specifications of the re-
ception-side personal computer.