7 - 2
No. 99MBC034B
7.4 NORMAL Mode
7.4.1 Overview of the NORMAL mode
•
The NORMAL mode is the factory-set default. Use it with-
out modification for general purposes.
•
Origin setting, tolerance setting, and presetting can be per-
formed individually for each CEL.
•
Tolerance judgment (for each CEL), BCD data, and seg-
ment (on selected CELs) output can be performed.
7.4.2 Tolerance judgment output
1) Function
Outputs the tolerance judgment result of each internal
counter (CEL1 to CEL6).
Measured value
<
Lower tolerance limit
-NG output
Lower tolerance limit
≤
Measured value
≤
Upper tolerance limit
GO output
Upper tolerance limit
<
Measured value
+NG output
2) Parameter setup
I/O function selection parameter: Set to NORMAL
I/O-type selection parameter:
Set to tolerance judgment
3) Pin assignment
1
19
18
36
No. Name
I/O
Functional description
1
COM
Common terminal of I/O circuit
2
COM
(connected to internal GND)
3
CEL1_-NG
OUT
Tolerance judgment result of CEL1
-NG
4
CEL1_GO
OUT
Tolerance judgment result of CEL1
GO
5
CEL1_+NG
OUT
Tolerance judgment result of CEL1
+NG
6
CEL1_NOM
OUT
Error signal of CEL1
*1
7
CEL2_-NG
OUT
Tolerance judgment result of CEL2
-NG
8
CEL2_GO
OUT
Tolerance judgment result of CEL2
GO
9
CEL2_+NG
OUT
Tolerance judgment result of CEL2
+NG
10
CEL2_NOM
OUT
Error signal of CEL2
*1
11
CEL3_-NG
OUT
Tolerance judgment result of CEL3
-NG
12
CEL3_GO
OUT
Tolerance judgment result of CEL3
GO
13
CEL3_+NG
OUT
Tolerance judgment result of CEL3
+NG
14
CEL3_NOM
OUT
Error signal of CEL3
*1
15
CEL4_-NG
OUT
Tolerance judgment result of CEL4
-NG
16
CEL4_GO
OUT
Tolerance judgment result of CEL4
GO
17
CEL4_+NG
OUT
Tolerance judgment result of CEL4
+NG
18
CEL4_NOM
OUT
Error signal of CEL4
*1