E-SOUM-PLUS Rev E
45
WARNING 1.4 Proper system grounding can only be ensured when an approved, hospital-grade receptacle and matching
power cord are used. To avoid the risk of electric shock, this equipment must only be connected to a supply
with protective earth. Install plug and receptacles as per local regulations before operating the unit. Power
cord, plug and receptacle should be examined to verify that they are in good working condition before
connecting the console. Never pull on the power cord to remove it from the receptacle.
WARNING 10.1 If a Mains Power fuse fails after replacement when the unit is reactivated, discontinue use of the device and
contact an authorized Misonix representative.
Lack of Irrigant
Symptoms
•
No spray from tip when ultrasound is engaged
•
No flush fluid available
•
Unexpected temperature rise at operative site
•
Unexpected temperature rise of handpiece
Alert Type
Alert Screen
Alert Action
None
None
None
Possible Cause
Ultrasound Mode
Corrective Action
1. Closed or empty fluid
bag
Set ultrasound to STANDBY.
Check fluid bag and tubing clamp. Replace fluid bag if neces-
sary.
2. Tubing not connected
Set ultrasound to STANDBY.
Check tubing connections.
Check mounting in pump head. Close pump cover until locked.
3. Tubing obstructed or
defective
Set ultrasound to STANDBY.
Check tubing for kinking, restrictions or leaks. Replace tubing if
necessary.
Check mounting in pump head. Close pump cover until locked.
4. Tubing installed in
reverse
Set ultrasound to STANDBY.
Open pump cover. Reposition the tubing in direction of flow.
Close pump cover until locked.
5. Tubing slides through
pump
Set ultrasound to STANDBY.
Open pump cover. Adjust the grip of the tubing by turning the
adjustment wheel underneath the front and back of the pump
assembly.
6. Pump defect
Set ultrasound to STANDBY.
Open pump cover. Check if pump rollers are rotating when
depressing footswitch. Replace console if they don’t.
Table 10.3 Troubleshooting – Insufficient Irrigation
WARNING 3.1
Tip and irrigant temperatures may exceed the tissue necrosis point if insufficient irrigation flow rates are
used. For hard tissue removal, set the irrigation flowrate to a setting no less than the comparable vibration
setting. For example, if the vibration setting is 7, a minimum flow setting of 70% should be used.
WARNING 7.1 Improper connection of the handpiece cable may present a shock hazard. Confirm that handpiece
connector is dry prior to plugging it in.
WARNING 7.2 Do not operate pump with pump cover in raised position. Rollers might pinch loose clothing or fingers.
Personal injuries may result. Cooling of the ultrasonic tip and irrigation at the treatment site will be inhibited.
For all other malfunctions please contact Misonix or a Misonix authorized representative for service.