E-SOUM-PLUS Rev E
28
7.
System Set-up
7.1.
Installation
Upon delivery perform a visual inspection of the shipping containers and all system components for obvious shipping
damage. Retain the shipping container and immediately notify the shipping carrier of any damages found.
CAUTION 7.1
All reusable system components like handpiece, probe front housing, counter wrench, and T-wrench and
autoclavable probe tips are supplied industrially cleaned, but NON-STERILE. All items intended for use in the
sterile field must be cleaned and sterilized as per the indicated instructions before first Clinical use and
before every subsequent Clinical use.
WARNING 12.1 Connecting the console to a power outlet with inadequate voltage or frequency may cause the unit to
malfunction or to create a shock or fire hazard. Confirm that the voltage selector switch on the console rear
panel is set to the local voltage setting and ensure that the correct fuses are being used. Refer to section
12.1 on instructions for fuse replacement. See section 12.1 for instructions on adjusting to local electrical
requirements.
Care should be taken to stay within the general operating conditions.
Operating Conditions
Operating conditions
•
Temperature 55-86°F (13-30°C)
•
Relative humidity 20-90% (non-condensing)
•
Standard atmospheric pressure
Table 7.1 Operating conditions
The console can be placed on an appropriate table or cart outside of the sterile field. Ensure that the pump head on the con-
sole right is installed. Refer to section 12.2 if the pump head is not yet installed.
Figure 7.1.1
Adjust the grip of the V-notches
Evenly adjust the V-notches to their fully
opened position by turning the adjust-
ment wheels.