E-SOUM-PLUS Rev E
31
7.4.
Console Set-up – Part II (Non-sterile)
Console Set-up | Part II
Connect Handpiece cable
Open pump cover
Attach cable connector receptacle on console front panel.
Align red dot on cable connector with red dot on front panel receptacle. Push cable connec-
tor into place.
Open the latch of the irrigation pump
The arrow on the pump housing indicates the direction of flow.
Insert tubing
Insert the soft silicone section by placing it over the pump rollers.
Secure tubing in V-notches
Verify that the tubing enters the pump from fluid container and exits to handpiece and in
direction of arrow on pump housing.
Place and hold the tubing’s silicone section in the V-notches on both pump sides.
Apply slight pressure to ensure that tubing rests within both V-notches.
Close pump cover
Close the latch of the irrigation pump until it locks.
Connect Tubing to
Connect IV-Spike to fluid container following standard operating room protocol.
Fluid Container
Prime Tubing
Adjust Settings
Irrigation tubing features vented IV-spike and is compatible with rigid bottles and flexible bags.
Check that ultrasound is in Standby Mode. Set Flow rate to 10.
Depress footswitch until fluid discharges at ultrasonic tip.
Set Amplitude and Flow to desired flow setting. Refer to section 4.1 for recommended set-
tings. Enable ultrasound.
Table 7.4 Console set-up - part II