9.4
Expected Life, Reusable Components
All handpiece components need to examined regularly, prior each use and be replaced if damaged.
The estimated sterilization life of handpiece components is listed below. All sterilization life estimates are approximate and
may be affected by rough handling, damage, wear due to vigorous cleaning, etc.
Estimated Sterilization Life
Item
Number Of Steam Sterilization Cycles
Handpiece with attached cable
~200 cycles
Probe covers
~300 cycles
Wrenches:
Handpiece/counter wrench and Probe wrench
~300 cycles
Probe
~200 cycles
Table 9.4 Estimated re-use life
NOTE 9.4
The reuse life given takes into account wear and tear due to cleaning and sterilization only. Damage or wear
caused by actual use in treatments will affect life of components.
9.5
Deviations From Decontamination, Cleaning And Sterilization Instructions
Misonix Inc. has validated all cleaning, disinfection and sterilization cycles given in this manual. It is highly recommended that
the procedures given in this manual for cleaning and sterilizing the SonicOne CLINIC system and related accessories be
followed. It is the responsibility of the user of this device or any accessories used with it to validate procedures for cleaning
and/or sterilization if they differ from the procedures as outlined in this manual.
Technical Assistance
Should the user wish further information or instructions regarding any aspect of cleaning or sterilizing procedures, please
contact Misonix Inc. or an Authorized Representative.
10.
Troubleshooting
The SonicOne CLINIC system provides both visual and audible alert signals when the system is not functioning properly.
Mechanical Limit Alert
Alert Type
Alert Screen
Alert Action
Mechanical Limit
Displays “LIMIT” alert located above amplitude setting
display.
Triggers a pulsed, audible indicator upon footswitch
activation.
Temporarily deactivates ultrasound and irrigation
functions.