Spirobank II
Rev.1.6
Page 17 of 35
User manual
Athmospheric pressure: 50kPa, 106 kPa
Operating conditions
Temperature: MIN + 10 °C, MAX + 40 °C;
Humidity: MIN 10% RH; MAX 95%RH
Athmospheric pressure: 50kPa, 106 kPa
Applied norms
IEC 60601-1:2005 + A1: 2012 (Electrical Safety)
IEC 60601-1-2:2015 (EMC)
ATS/ERS Guidelines: 2005, 2019
ISO 26782: 2009
ISO 23747: 2015
EN ISO 14971: 2019
ISO 10993-1: 2018
2011/65/UE Directive
EN ISO 15223:2016
EN IEC 60601-1-6: 2010+Amd2013
ISO 80601-2-61: 2017
IEC 60601-1-8: 2006
Essential performances (according to
IEC60601-1:
2005 + A1: 2012)
Error of displayed numeric value: Flow measurement percentage error < ± 5%
Measure of the oximetry parameters with accuracy defined in table on § 1.6.2
Emission limits
CISPR 11 Group 1 Class B
Electrostatic discharge protection
8kV contact, 15kV air
Magnetic field immunity
30 A/m
Radio Frequency Immunity
3V/m @ 80-2700 MHz
MIR will make available on request circuit diagrams, component part lists, descriptions, calibration instructions, or other information
that will assist service personnel to repair those part of the device that are designated by MIR as repairable by service personnel.
2
FUNCTIONING OF THE Spirobank II
2.1
Switch on and switch off the device
To switch on the
Spirobank II
push
The first screen shows the manufacturer logo, information of date and hour
set on the device. If no key are touched, after a few seconds the device shows
the main screen.
The second screen shows the information as in the image beside. He key
allows to visualize the service menu; with these items it is possible to configure
the device in the correct way.
If any keys are touched, then the device goes to the main screen.
To switch off the device push
.
WARNING