4. HANDLING PRECAUTIONS
Sensor is not explosive protected.
•
Do not use damaged sensor. It must be repaired only by personnel authorized by
manufacturer.
•
Keep sensor out of contact with aggressive substances e.g. acidic environments, which
can react with metals, as well as solvents, which may affect polymeric materials.
•
Diffusion holes of sensor should be protected against ingress of dust and sprayed
materials.
•
Sensor is not intended to measure target gas concentration contained in fluids.
•
Maximum allowable pressure on the central part of sensor reflecting cover or on sensor
side surface – 2 MPa, on sensor upper edge – 100 MPa.
•
Sensor updates information about concentration every 1.28 ± 0.065 second. Sending any
command more often than one time per second (1 Hz) may reduce sensor accuracy.
•
Correct measurement is provided when ambient temperature changes not faster than 0.6
°C/min.
•
Covering diffusion holes of sensor increases its response time t(90).
•
When operating sensor, observe conditions stated in Table 2 and Table 4.
•
When dealing with cylinder containing gas mixture under pressure, it is necessary to follow
the safety regulations. Dumping of CGM into the work area is not allowed.
•
There is no risk of pollution and negative impact on human health. Sensor does not contain
any harmful substances that may be released during its normal operation.
•
The
sensor
is
manufactured
in
conformity
with
the
declaration
RoHS 2 Compliant Directive 2011/65/EU.
•