THE SOLE PROPERTY OF SIA MIPEX. ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED.
Table 6. Basic sensor readings variability
5
Calibration
gas
Readings variability within a temperature
range
Additional variabil-
ity due to pressure
Additional variabil-
ity due to humidity
CO
2
± 0.05% vol. or ± 5% of readings (whichever
is greater) within the range of +20…+25 °C;
± 0.1% vol. or ± 40%
of readings
(whichever is
greater)
at 100 kPa
(tested at 80 kPa, 100
kPa, 120 kPa)
± 0.1% vol. or ± 15%
of readings
(whichever is
greater)
at 40 °C
(tested at 20% RH,
50% RH, 90% RH)
± 0.1% vol. or ± 10% of readings (whichever
is greater) within the range of -10…+20 °C
and +25…+40 °C;
± 0.2% vol. or ± 20% of readings (whichever
is greater) within the range of -20…-10 °C
and +40…+50 °C.
5
The table shows basic variability of MIPEX-02 sensor readings. For each sensor modification, readings
variability stated in Table 6 is provided within a temperature range determined by a sensor RX-code (see Table
5).