72
Data Processor
Prepation
Basic Setting
The instrument has the following nine basic settings.
Item
1) Printer
(Automatic
printing)
2) Printing color
spaces
3) Data
Protection
4) Number of
measurement
for auto. average
5) Illuminant
6) Back light
7) Buzzer
8) Displayed
color limit
9) Remote mode
Settings
OFF/ON
OFF/ON
OFF/ON
1 - 30
measurements
C/D
65
OFF/ON
OFF/ON
XYZ, Yxy,
L*a*b*,
Hunter Lab,
L*C*h, Munsell
(illuminant C
only), CMC (l:c),
CIE1994, Lab99,
LCh99, CIE2000,
CIE WI/Tw
(illuminant D
65
only), WI ASTM
E313 (illuminant
C only), YI ASTM
D1925 (illuminant
C only), YI ASTM
E313 (illuminant
C only)
*User index
Default
ON
OFF
OFF
1
measurement
C
OFF
ON
XYZ, Yxy,
L*a*b*,
Hunter Lab,
L*C*h,
Munsell
Description
This sets whether or not to automatically print data
after measurement.
Print/Paper Feed key still works, regardless of this
setting.
This sets whether or not to print all color spaces se-
lected in the display limit.
(This does not include statistical operations.)
This sets whether or not to protect existing data in the
event that the number of stored data sets exceeds 2000.
•
If data protection is OFF, old data is overwritten.
•
If data protection is ON, the 2001st data is pre-
vented.
This sets the number of measurements for automatic
average.
It can be set to between 1 and 30 measurements.
This sets the illuminant to C or D
65
.
The illuminant cannot be changed during measure-
ment. Set illuminant again to change an illuminant
after initial setting (See P.76). Some color spaces may
not be able to be displayed depending on the
illuminant setting. (See P.79 “Color Space and Chang-
ing the Display”.) An error message is displayed and
connection is prevented if different illuminants are
set for the measuring head and the data processor.
This sets the back light ON and OFF.
The back light automatically goes OFF 30 seconds
after key operation, even if it is set to ON.
This sets the buzzer ON and OFF.
The following buzzer sounds are emitted when set to
ON.
Key reception, error, judgment, settings complete.
Select the color space to be displayed.
•
The color space currently displayed in the measure-
ment screen cannot be turned OFF.
•
It is not possible to turn OFF all color spaces.
•
Some color spaces cannot be selected depending
on the illuminant setting.
This mode is for data output to connected PC.
*When set to the measuring head (See P.127 “User Index”.)
Summary of Contents for CR-400 -
Page 23: ...21 Chapter 1 Measuring Head 1 This chapter describes how to use the measuring head alone...
Page 54: ...52...
Page 143: ...141 Chapter 4 Description...
Page 156: ...154 Description Dimensions Measuring Head Unit mm CR 400 Head 217 63 102 40 2 32...
Page 157: ...155 Description Unit mm CR 410 Head 67 5 244 63 102 32...
Page 158: ...156 Description Data Processor Unit mm DP 400 Dimensions 100 255 73...