7-3. Screen Description
Name
Description
Start Calibration
Starts calibration.
ZERO
AD Counts
Shows the AD count value with the torque transducer in the initial torque state.
kHz
Shows the frequency conversion value with the torque transducer in the initial
torque state.
Register button
Tentatively registers the torque transducer initial torque state.
<<, <, >, >>
Adjust voltage output in the initial torque state.
+SPAN
AD Counts
Shows the AD count value with the torque transducer in the span torque state.
kHz
Shows the frequency conversion value with the torque transducer in the span
torque state.
Register
Tentatively registers the state with span torque applied to the torque transducer.
<<, <, >, >>
Adjust voltage output in the span torque state.
–SPAN
AD Counts
Shows the AD count value with the torque transducer in the minus torque state.
kHz
Shows the frequency conversion value with the torque transducer in the minus
torque state.
Register
Registers the state with minus torque applied to the torque transducer.
<<, <, >, >>
Adjust voltage output in the minus torque state.
Register
Registers calibration.
Stop Calibration
Stops calibration.
Close
Returns to the menu screen.
Status:
Shows the status.
Offline: Communication stopped
Online: Communication in progress (calibration in progress)
Torque AD Value
AD Counts
Shows the current TMHSA internal AD value.
Torque AD Value
kHz
Shows the current TMHSA input frequency.