14 
 

7-4.  Calibration Procedure 

 

(1) Step 1: Start calibration   

Click [Start Calibration]. Calibration begins. 

 

 

(2) Step 2: Zero calibration   

With the torque transducer in the initial torque state, click [Register] in ZERO.   
The voltage output becomes approximately 0 V. 

 

Adjust the voltage output using [<<] (coarse negative adjustment), [<] (fine negative adjustment), 
[>>] (coarse positive adjustment), and [>] (fine positive adjustment). 
 

(3) Step 3: Plus span calibration   

With span torque applied to the torque transducer, click [Register] in +Span.   
Voltage output becomes approximately 10 V. 

 

Adjust the voltage output using [<<] (coarse negative adjustment), [<] (fine negative adjustment), 
[>>] (coarse positive adjustment), and [>] (fine positive adjustment). 
 

(4) Step 4: Minus span calibration   

With minus torque applied to the torque transducer, click [Register] in -Span.   
Voltage output becomes approximately -10V. 

 

Adjust the voltage output using [<<] (coarse negative adjustment), [<] (fine negative adjustment), 
[>>] (coarse positive adjustment), and [>] (fine positive adjustment). 
 
 

 

Summary of Contents for TMHSAD-01

Page 1: ...TORQUE TRANSDUCER CALIBRATION SOFTWARE TMHSAD 01 Instruction Manual EN294 1799 ...

Page 2: ......

Page 3: ...in the preparation of this instruction manual it may not be possible to correct errors or omissions immediately This instruction manual is copyrighted by our company It may not be reproduced in part or in whole without our express permission Scope of Software License The software refers to the TMHSAD 01 software and its accompanying documentation including the instruction manual A copy may not be ...

Page 4: ...oid the actions described here at all times Caution This indicates circumstances in which incorrect handling may result in injury to users or damage to property This indicates operating or procedural precautions or restrictions Always read the details included here to avoid malfunctioning Safety Precautions Please be sure to read this manual before attempting to use the equipment 1 Precautions Cau...

Page 5: ...III Revision History Date Manual No Revision Reason Details April 2018 DRW No EN294 1799 First edition ...

Page 6: ...on 12 7 1 What Is Calibration 12 7 2 Calibration Procedure 12 7 3 Screen Description 13 7 4 Calibration Procedure 14 8 Electrical Calibration 16 8 1 What Is Electrical Calibration 16 8 2 Screen Description 16 8 3 Calibration Procedure 18 8 4 Adjusting Voltage Output 19 9 Parameter Settings 20 9 1 Screen Description 20 9 2 Reading Setting Parameters 21 9 3 Writing Setting Parameters 22 9 4 Setting ...

Page 7: ...V 12 Uninstalling 31 ...

Page 8: ......

Page 9: ...inimum working memory 2 GB or more Hard disk drive Free space 1 GB or more CD ROM drive DVD drive also acceptable Required for setup Available USB ports 1 Required for USB cable communication Minimum resolution 1 280 1 024 3 Usage Precautions Never turn off the TMHSA power or disconnect the USB cable while reading or writing setting parameters to or from the TMHSA If this happens do not leave the ...

Page 10: ...C to be used as follows 1 Insert the TMHSAD 01 setup disc CD ROM into the PC 2 The AutoPlay screen appears If this does not happen automatically open the CD ROM in Windows Explorer or File Explorer and double click TMHSAD 01_English msi in the Setup Setup_English folder 3 The Setup Wizard screen appears automatically after the Windows Installer Dialog is displayed Click Next ...

Page 11: ...ftware and then click Next To change the installation destination folder To install to a different location click Browse Once the Browse Folder dialog appears specify the folder in which to install the software and then click OK 5 The Confirm Installation screen appears Click Next Installation starts ...

Page 12: ...4 6 Installing TMHSAD 01 appears Wait for installation to be completed The screen changes automatically once installation is complete 7 Installation Complete appears Click Close ...

Page 13: ...ia USB Install the driver as follows The USB drivers are included on the CD ROM provided 4 2 1 Driver Installation Procedure 1 Insert the CD ROM provided in the CD ROM drive 2 Double click USB_Driver CDM21226_Setup exe on the CD ROM provided to start installation 3 The FTDI CDM Drivers screen appears Click Extract ...

Page 14: ...6 4 The device driver installation wizard screen appears Click Next 5 Select to accept the license agreement and click Next ...

Page 15: ... device driver installation wizard complete screen appears Click Finish Installation ends Screen while installation is in progress Screen when installation ends 7 The TMHSA will be recognized automatically when connected to the PC via USB and is ready for use ...

Page 16: ...meters Check the version and read write setting values For details see 9 Parameter Settings Actual Torque Calibration Calibrate under conditions of actual torque For details see 7 Calibration Electrical Calibration Enter the AD values to perform calibration For details see 8 Electrical Calibration Communication settings Set up communication with the TMHSA For details see 6 Communication Settings L...

Page 17: ...COM port number has probably changed Check the COM port and confirm that it matches the port number in the communication settings 6 1 Checking the COM Port Check the COM port for the TMHSA connection 1 Right click the desktop Start menu and select Device Manager in the displayed menu In case of Windows 7 click the desktop Start menu and right click Computer and select Management Click Device Manag...

Page 18: ...in in Device Manager and check the COM port number Example COM4 is used in the following case 6 2 Setting the COM Port 1 On the menu screen click Communication settings 2 Select the COM port for the TMHSA connection and click Save ...

Page 19: ...m normal communication Communication is normal if 200 appears in the RXD box The following message box is displayed if communication was not successful See the previous instructions and recheck the communication setting details ...

Page 20: ...er 7 2 Calibration Procedure Switch to calibration mode Set the torque transducer to the initial torque state and perform zero calibration With span torque applied to the torque transducer register the plus span point With minus torque applied to the torque transducer register the minus span point Connect the torque transducer to the PC Turn on and wait for approximately 10 minutes for the torque ...

Page 21: ...egister Tentatively registers the state with span torque applied to the torque transducer Adjust voltage output in the span torque state SPAN AD Counts Shows the AD count value with the torque transducer in the minus torque state kHz Shows the frequency conversion value with the torque transducer in the minus torque state Register Registers the state with minus torque applied to the torque transdu...

Page 22: ...justment 3 Step 3 Plus span calibration With span torque applied to the torque transducer click Register in Span Voltage output becomes approximately 10 V Adjust the voltage output using coarse negative adjustment fine negative adjustment coarse positive adjustment and fine positive adjustment 4 Step 4 Minus span calibration With minus torque applied to the torque transducer click Register in Span...

Page 23: ...15 5 Step 5 Registration Click Register to register calibration Setting values are only tentatively registered until you click Register Setting values are not saved if you click Stop Calibration midway ...

Page 24: ...e initial torque state SPAN AD Counts Shows the AD count value with the torque transducer in the span torque state Enter a value directly to change the setting value kHz Shows the frequency conversion value with the torque transducer in the span torque state SPAN AD Counts Shows the AD count value with the torque transducer in the minus torque state Enter a value directly to change the setting val...

Page 25: ...transducer calibration AD count values and configuration information VOUT Calibration Start Calibration Starts voltage output calibration Stop Calibration Stops voltage output calibration ZERO SPAN SPAN Adjust voltage output when in the selected state Register Registers calibration Close Returns to the menu screen Status Shows the status Offline Communication stopped Online Communication in progre...

Page 26: ...ollowing message appears Click Yes Calibration data is displayed 3 Step 3 Adjust the torque transducer AD count value and Configuration settings as needed 4 Step 4 Click Write The torque transducer calibration AD count value and configuration information is written ...

Page 27: ...tput adjustment begins 2 Step 2 Select an item for voltage adjustment ZERO SPAN or SPAN Voltage is output in the selected state 3 Step 3 Use and to adjust the voltage of the selected item 4 Step 4 After the adjustment click Register Setting values are only tentatively registered until you click Register Setting values are not saved if you stop midway ...

Page 28: ...ws relevant setting information Calibration SPAN AD Value item Shows relevant setting information Calibration SPAN AD Value item Shows relevant setting information Moving Average Shows relevant setting information Polarity item Shows relevant setting information DAC Filter item Shows relevant setting information Read button Reads information Write button Writes information Save button Saves the in...

Page 29: ... 2 Reading Setting Parameters This section describes the procedure for reading TMHSA setting parameters 1 Click Read on the screen 2 Click Yes in the message box 3 The TMHSA setting parameters are displayed ...

Page 30: ...once 2 The TMHSA setting parameters are displayed in Read and Write Enter a Write value as desired for the setting parameter to change For details of the input ranges and selection items for each item see 9 4 Setting Parameter List The cells for items for which overwriting is prohibited are grayed out input and selection disabled 3 Click Write on the screen 4 Click Yes in the message box The infor...

Page 31: ...span AD count value Calibration SPAN AD Value 0 262143 Calibration minus span AD count value Moving Average None 2 times 4 times 8 times 16 times 32 times 64 times 128 times 256 times 512 times Polarity output CCW output CW Counterclockwise plus voltage output Clockwise plus voltage output DAC Filter 1 kHz 500 Hz 300 Hz 100 Hz 50 Hz 30 Hz 10 Hz 1 Hz indicates default settings ...

Page 32: ...justed 1 Click Calibration The button changes to Editing Calibration items can be set in this Editing mode 2 Adjust the values of the calibration items to set 3 Click Start after changing calibration items 4 The following message appears Click Yes Calibration item writing is complete ...

Page 33: ...ormat 1 Click Save 2 The following message appears Click Yes Calibration item writing is complete 3 The Please select a folder to save file dialog is displayed Specify the save destination and file name and save the results If Cancel is clicked data will not be saved ...

Page 34: ... 1 Click Load 2 The following message appears Click Yes Calibration item loading is complete 3 A dialog is displayed to select a file to open Specify where to load the file from and the file name and click Open 4 Values loaded from the file are shown in the Write items ...

Page 35: ...hows the wave height voltage value Hi wave height value Red when the wave height is high Mid wave height value Red when the wave height is moderate Lo wave height value Red when the wave height is low A Z Status display Shows the A Z status A Z ON button Activates A Z A Z OFF button Deactivates A Z CHECK Status display Shows the check status CHECK button Switches between checking on off Close Retu...

Page 36: ...MHSA 10 4 Torque AD Value Shows the torque AD count value and the corresponding value converted to frequency 10 5 Wave Height value Shows the voltage value and status of the torque meter light level Screen Details Normal torque meter light level Low torque meter light level The torque meter light level is too low to guarantee torque measurements ...

Page 37: ...A Z Click A Z ON to activate A Z ON status Click A Z OFF to deactivate A Z OFF status 10 7 CHECK Click CHECK to switch between checking on off Outputs approximately 8 0 V of analog voltage ON status OFF status ...

Page 38: ... save the contents of language selection To apply the setting of language selection please quit the calibration software and start it again Close Returns to the menu screen If you do not save the settings the change of language selection will not be effective ...

Page 39: ...ftware 1 Right click the desktop Start menu and select Apps and features in the displayed menu In case of Windows 7 select Programs and functions in the Control panel 2 Select TMHSAD 01 and run Uninstall 3 Uninstalling is complete once the window below disappears ...

Page 40: ......

Page 41: ......

Page 42: ...e Product Sales Management 1 1 1 Katase Fujisawa shi Kanagawa ken 251 8531 Japan Tel 81 466 23 2681 Fax 81 466 22 7191 Sensing Device Business Unit FUJISAWA PLANT 1 1 1 Katase Fujisawa shi Kanagawa ken 251 8531 Japan Tel 81 466 22 7151 Fax 81 466 22 1701 KARUIZAWA PLANT 4106 73 Miyota Miyota machi Kitasaku gun Nagano ken 389 0293 Japan Tel 81 267 31 1309 Fax 81 267 31 1353 HOMEPAGE ADDRESS http ww...

Reviews: