21
5
CPU normal operation
Notify that CC
−
LINK interface is operating normally. Reverse the status of ON/OFF in 0.5
seconds.
6
Decimal point position 1, 2, 3 or 4
Notify the decimal point position of the load value by the binary value of three points. This
output is updated by turning on the power supply, and initialed data set request flag (RY0079).
0
:
No decimal point
1
:
10
0
digit
2
:
10
2
digit
3
:
10
3
digit
4
:
10
4
digit
7
S0
〜
S2
Notify the condition of S0
〜
S2. The same condition with S0
〜
S2 of the indicator
8
Holding
Notified whether the load value is holding.
ON
:
Holding
OFF
:
Free running
9
Abnormal load value
Notifies when the load value is “OL” or “
−
OL”.
ON
:
When abnormality occurs
OFF
:
Normal
10
Initialed data set completion flag
Notify the end of initialization when there is a request with initialed data set request flag
(RY0079).
ON
:
In the completion of set
OFF
:
Normal
11
Error condition flag
Notify when the error occurs in the indicator. After the error is released, it is reset with error
reset request flag (RY007A).
ON
:
In the occurrence of error
OFF
:
Normal
12
Remote ready
Notified to be able to complete initialization and to communicate.
ON
:
Possible to communicate
OFF
:
In the initialization