17
7
. Specifications
7
−
1. Specifications
Power supply voltage to rotor AC20.0 Vrms
Square wave 11.2 kHz
±
0.2 kHz
Applicable transducer
Optical transformer method torque transducer
Input range
5 kHz to 15 kHz
Output
±
10 V
Output load resistance
Load resistance 2 k
Ω
or more
Load of output capacitance 0.1
μ
F or less
Zero adjustment range
±
15 %R.O.
Non
−
linearity
0.01 %F.S.
Temperature coefficient
Zero point
±
0.025 %R.O./10
℃
Sensitivity
±
0.025 %R.O./10
℃
CHECK
Approx.80 %R.O
Accuracy
±
0.03 %
Frequency response range
DC to 1 kHz (Filter
:
W/B)
(Changeable to 1 Hz, 10 Hz, 30 Hz, 50 Hz, 100 Hz, 300 Hz, 500 Hz or 1 kHz)
However, 10 Hz to 1 kHz
:+
1 dB,
−
3 dB
±
1 dB, 1 Hz
:+
1 dB,
−
3 dB
±
3 dB
Display section
Display of output voltage 0 to
±
10.000 digital display (Green LED)
Sampling
Approx. 4 times/s
Remote function
CHECK
ALARM function
ALARM output
7
−
2. General specifications
Operating temperature/humidity range
0
℃
to 40
℃
85 %RH or less (Non condensing.)
Power supply
Power supply voltage
AC85 V to AC132 V
Power supply frequency 50/60 Hz
Power consumption
Approx. 100 VA (at AC100 V)
Insulation resistance
Between power line and case
DC500 V, 100 M
Ω
or more
Between power line and case
AC1 500 V, 1 min period
Outline dimension(W
×
H
×
D) 42 mm
×
176 mm
×
284.4 mm (Excludes protruding parts.)
Weight
Approx.2.5 kg
7
−
3. Standard specifications at the shipment
SPAN
10 V output at rated output from torque transducer
Frequency response range
1 kHz
7
−
4. Accessories
Instruction manual
1 piece
Fuze(5 A)
1 piece
Output cable(2 m)
1 piece
Dust filter
10 pieces