22
●
In confirming the voltage between A and C, please set the connection range of the measuring instrument such as tester to DC
・
V.
●
In confirming the voltage between B and D, please set the connection of the measuring instrument to DC
・
mV.
Remove the cable of strain gage applied
transducer from the terminal
Shorten between B & D, and A & G
Voltage/Current output
are stable
Same condition
Start measurement
4
YES
NO
NO
YES
NO
Inform Minebea about the contents of
failure and situation at site in details.
Confirm various connection according to the paragraph 2.
Especially, please confirm whether there is influence from the
noise source such as inverters.
Inform Minebea about the contents of
failure and situation at site in details.
YES
Voltage between A & C
is the same as specified, and
stable.