24
8-4. Options
8-4-1. Auto zero
P/N
CSA522B-P99
Input range of auto zero
within
±
2.4 mV/V
auto zero range
within
±
10 V
Required time for auto zero
within approx.1 s
Accuracy of auto zero
within
±
5 mV
Back-up time
Approx. 10 years (with the usage of Lithium battery)
Temperature
coefficient
Zero point
±
0.005 %F.S./
℃
Sensitivity
±
0.005 %F.S./
℃
8-4-2. Power supply voltage
P/N
CSA522B-P61
Power supply voltage
AC110 V (Permissible variable range AC99 V to AC127 V)
Power supply frequency
50/60 Hz
Power consumption
Approx. 15 VA
P/N
CSA522B-P63
Power supply voltage
AC200 V (Permissible variable range AC180 V to AC220 V)
Power supply frequency
50/60 Hz
Power consumption
Approx. 15 VA
P/N
CSA522B-P64
Power supply voltage
AC220 V (Permissible variable range AC198 V to AC253 V)
Power supply frequency
50/60 Hz
Power consumption
Approx. 15 VA
8-5. Standard specification at the shipmennt
Bridge power supply
DC10 V
Sensitivity adjustment
1 000 times (DC10 V output at 1.0 mV/V input)
Range of frequency responce
30 Hz (-12 dB/oct) Bessel type, and 25 kHz (-3 dB)