4. CE conformity standard
This instrument has suited the following standard.
EN61326−1
:
2006
“Electrical equipment for measurement, control, and laboratory use − EMC requirements”
“Immunity test requirements for equipment intended for use in industrial locations”
EN61010−1
:
2001
“
Safety requirements for electrical equipment for measurement, control and laboratory use”
The operating condition to conform this standard is as follows :
Caution
4-1. Place of installation
Please set up this instrument in the shielded case or control panel where EMC measures are
given.
4-2. Wiring
①
Shield processing
Please make sure to shield all the signal cable by using the shielded cable or using the conduit
piping including the storage case and control panel.
Please make sure to shield the power cable by using the conduit piping including the storage case
and control panel.
②
Grounding
Please make sure to apply the grounding through the case and control panel where EMC
measures are given with protective earth terminal.
Please observe the following conditions strictly when this instrument suits the above
−
mentioned standard.
If neglected, there is a possibility of not suiting the above
−
mentioned standard.