Designation
Description
Abbr.
D1
Unit
Output impedance
between measuring terminals
R
O
1200 ±2
Ω
Insulation impedance between measuring circuit and housing,
U
DC
= 100 V
R
IS
>5000
M
Ω
Recommended sup-
ply voltage
to hold the speciied performance
B
u
4
24
V
Max. supply voltage
permissible for continuous operation wi-
thout damage
U
max
32
V
Nominal ambient
temp. range
to hold the speciied performance
B
T
-10
+85
°C
Usable ambient temp.
range
permissible for continuous operation wi-
thout damage
B
Tu
-30
+95
°C
Storage temperature
range
without electrical and mechanical stress
B
Ti
-40
+95
°C
Permissible eccentri-
city
permissible displacement from nominal
load line
S
ex
3.0
mm
Vibration resistance
resistance against oscillations
(IEC 60068-2-6-Fc)
20 g, 100 h,
10
150 Hz
Barometric pressure
inluence
inluence of barometric pressure on output PK
Smin
≤12
g/kPa
Nominal delection
elastic deformation under maximum capa-
city
S
nom
<0.1
mm
Deinitions acc. to VDI/VDE 2637
The technical data given are intended solely as a product description and should not be
interpreted as guaranteed properties in the legal sense.
Compact Compression Load Cell PR 6211
4 Speciications
EN-12
Minebea Intec