![Minebea Intec IS16EDE-H Installation And Operating Instructions Manual Download Page 12](http://html.mh-extra.com/html/minebea-intec/is16ede-h/is16ede-h_installation-and-operating-instructions-manual_1801139012.webp)
12
Installation and Operating Instructions IS16/34/64EDE
Specifications Charts
12
IS34EDE-H0CE (Type of weighing instrument: SART BF BF)
verifiable for use in legal metrology.
Specifications group
Spec. gp.1
Spec. gp.2*
Spec. gp.3
Spec. gp.4
Spec. gp.5
Spec. gp.6
(code):
(10 3 1)
(10 3 2)
(10 3 3)
(10 3 4)
(10 3 5)
(10 3 6)
Maximum capacity
kg
34
34
26.1
34
26.1
34
Readability
g
0.1 0 to 6kg: 0.1
0 to 6kg: 0.1 1
1
0 to 6 kg 0.1
6 to 15kg: 0.2
6 to 15kg: 0.2
6 to 34 kg 1
15 to 34kg: 0.5 15 to 26.1kg: 0.5
Weighing range
Standard
Configurable Configurable
Configurable
Configurable
Configurable
Zero tracking±2%
kg
±0.68
±0.68
±0.52
±0.68
±0.52
±0.68
Initial zero-setting range
kg
±1
±1
±1
±1
±1
±1
Preload
1
)
kg
4
4
12
4
12
4
isoTEST function
2
)
yes
yes
yes
yes
yes
yes
Metrological specifications
Accuracy class
kg
K
K
K
K
K
K
Minimum capacity
acc. to CD
3
)
g
5
5
5
50
50
5
Minimum capacity
acc. to AWC
4
)
g
150
150
150
150
150
150
Verification scale interval g
1
1
1
1
1
1
Temperature range
°C
10 to +30
+10 to +30 +10 to +30
+10 to +30
+10 to +30
+10 to +30
1
) No reduction in the maximum capacity up to this preload; higher preloads reduce the maximum capacity in 1-kg steps. You cannot set a
preload > approx. 31 kg.
2
) The isoTEST function cannot be selected for preloads > approx. 21 kg
3
) CD = Council Directive 2009/23/EC on non-automatic weighing instruments used in the EU and the Signatories of the Agreement on the
European Economic Area
4
) AWC = average weight control based on the requirements of the German Regulations for Prepackaged Products
5
) Not allowed for use in legal metrology (legal for trade). A 10-kg calibration weight is required for external adjustment.
* = factory setting
IS34EDE-H0CE(Type of weighing instrument:
SART BF BF)verifiable for use in legal metrology.
Specifications group
Spec. gp.6
Spec. gp.7
Spec. gp.8
Spec. gp.9
(code):
(10 3 6)
(10 3 7)
(10 3 8)
(10 3 9)
Maximum capacity
kg
34
26.1
34
26.1
Readability
g
0 to 6 kg: 2 0 to 6 kg: 2 5
5
6 to 15 kg: 5 6 to 15 kg: 5
15 to 34 kg: 10 15 to 34 kg:10
Weighing range
Configurable
Configurable Configurable
Configurable
Zero tracking±2%
kg
±0.68
±0.52
±0.68
±0.52
Initial zero-setting range
kg
±1
±1
±1
±1
Preload
1
)
kg
4
12
4
12
isoTEST function
2
)
yes
yes
yes
yes
Metrological specifications
Accuracy class
kg
l
l
l
l
Minimum capacity acc. to CD
3
) g
40
40
100
100
Minimum capacity acc. to AWC
4
) g
500
500
2500
2500
Verification scale interval g
2/5/10
2/5/10
5
5
Temperature range
°C
+10 to +40
+10 to +40 +10 to +40
+10 to +40
1
) No reduction in the maximum capacity up to this preload; higher preloads reduce the maximum capacity in 1-kg steps. You cannot set a
preload > approx.31 kg.
2
) The isoTEST function cannot be selected for preloads > approx. 21 kg
3
) CD = Council Directive 2009/23/EC on non-automatic weighing instruments used in the EU and the Signatories of the Agreement on the
European Economic Area
4
) AWC = average weight control based on the requirements of the German Regulations for Prepackaged Products
5
) Not allowed for use in legal metrology (legal for trade). A 10-kg calibration weight is required for external adjustment.
* = factory setting
Summary of Contents for IS16EDE-H
Page 20: ...20 Installation and Operating Instructions IS16 34 64EDE Declaration of Conformity 1 6 ...
Page 21: ...Installation and Operating Instructions IS16 34 64EDE 21 Declaration of Conformity 2 6 ...
Page 27: ...Installation and Operating Instructions IS16 34 64EDE 27 Certificate ...
Page 29: ...Installation and Operating Instructions IS16 34 64EDE 29 Certificate ...
Page 30: ...30 Installation and Operating Instructions IS16 34 64EDE Plates and Markings 26 ...
Page 31: ...Installation and Operating Instructions IS16 34 64EDE 31 Plates and Marking 27 ...
Page 33: ...Installation and Operating Instructions IS16 34 64EDE 33 Plates and Marking PPIS030511e ...
Page 39: ...Installation and Operating Instructions IS16 34 64EDE 39 ...