Designation
Description
Abbr. D1, D1E
C3, C3E
C6, C6E
Unit
Insulation voltage
between circuit and
housing (../E versions on-
ly)
500
V
Recommended sup-
ply voltage
to hold the speciied per-
formance
B
u
4
24
V
permissible for conti-
nuous operation without
damage
U
max
32
V
Max. supply voltage
../E versions:
U
max
25
V
Nominal ambient
temp. range
to hold the speciied per-
formance
B
T
-10
+40
°C
Usable ambient
temp. range
permissible for conti-
nuous operation without
damage
B
Tu
-40
+95
°C
Storage temperature
range
without electrical and
mechanical stress
B
Ti
-40
+95
°C
permissible displacement
from nominal load line at
the head of the load cell
E
max
≤10 t
S
ex
10
mm
Permissible eccentri-
city
E
max
>10 t
S
ex
5
mm
Vibration resistance resistance against oscilla-
tions (IEC 60068-2-6-Fc)
20 g, 100 h, 10
150 Hz
inluence of barometric
pressure on output
up to E
max
= 3 t
PK
Smin
200
g/kPa
E
max
= 5
10 t
PK
Smin
330
g/kPa
Barometric pressure
inluence
E
max
= 20
75 t
PK
Smin
420
g/kPa
elastic deformation un-
der maximum capacity
up to E
max
= 5 t
s
nom
≤0.3
mm
E
max
= 10
20 t
s
nom
≤0.5
mm
E
max
= 30
50 t
s
nom
≤0.8
mm
Nominal delection
E
max
= 60
75 t
s
nom
≤1.2
mm
Deinitions acc. to VDI/VDE 2637
The technical data given are intended solely as a product description and should not be
interpreted as guaranteed properties in the legal sense.
Compression load cell Inteco® PR 6203
4 Speciications
EN-14
Minebea Intec