48
Service Manual Combics
inf 07
inf 08
Inf 09
inf 22
inf 23
inf 29
30th Inf 31
Inf 54
Inf 55
inf 71
inf 72
Cannot store the current weight value
None
(e.g. transaction counter maximum reached)
Display
Cause
Solution
Operating Design | Error Description Display Combics | 1 | 2 | 3
Function not allowed in scales not verified
for legal use
The load on the scale is too heavy to enable
zero positioning
Taring is not possible when the scale gross
weight is zero
Error in storing reference value,
Weight is too low
Error in initializing an application
Minimum load not reached
XBPI mode activated or interface not OK
A/D converter control below minimum limit,
the A/C converter provides no or a much
too low measured result.
A/D converter control over maximum limits;
the A/C converter provides a much too high
measured result.
Cannot store the current weight value
(or entry) (e.g. control limits too low or
too high)
Contact your local Service Center for information
on changing settings
Check whether in your configuration “zero at
power on” (1.12) is set
Zero the scale
Place a heavier weight on the platform
Contact your local Service Center
Define a lower value for the minimum load
(under Application, menu item 3.6)
Change menu, execute "Close" function,
Cable/Interface not OK
Contact your local Service Center
Check the weighing system, digital PCB and the
connection between them. If required, replace
the PCB or repair the weighing system.
Check the weighing system, digital PCB and the
connection between them. If required, replace
the digital PCB.
None