12
Service Manual Combics
Verification access switch (or access switch)
This switch must be open to use the Service software or carry out ADC configuration
(e.g. verification). If closed, the error message “ACCESS-SWITCH LOCKED” appears in the
Service program and the indicator unit displays "L" for underweight or "H" for overload.
ADC configuration is no longer possible.
The ± 2% window is active for required adjustments (characteristic value, linearity).
CAIS3, CAISL3
Display (graphic-capable dot-matrix display),
On/Off key.
General function keys:
(Scale) When connecting a second weighing
platform, the display switches between
the platforms, Zero, Tare, Switch function,
Adjustment/Calibration, Print/Data output,
(x10) Toggle unit between normal and 10-fold
higher resolution, (B/G) Net/Gross value, Clear
function, Reference value, Apply value, Switch
display, Toggle between available applications,
Setup, Memory, Decimal keypad
Handling and Operation
A
Closed
Open
Summary of Contents for CAIS3