13.  The “

About”

 screen contains system information such as 

Model,

 

Firmware, Bootloader, and Serial #.

14.  Once Settings have been adjusted return to 

Local Mode

 main

 

screen by using the “

Back

” tab. When on this screen, the “

Pre-Acq

” 

 

tab can be selected to move to the 

Pre-Acquisition screen

.

15.  The pre-acquisition screen shows the real time signals that are being read by the device. The    

 

channels can be cycled through by using the 

left and right arrow keys

. It is a good Idea to check 

 

the signals in the acquisition screen before acquiring data. Note that the 

“Back”

 tab can be  

 

 

selected to return to previous screens. 

16. Once 

Pre-Acquisition

 has been reviewed, data collection can 

 

begin by selecting the 

“Acquire”

 tab.

17.  At this point, data is being recorded to the SD card. 

Do not remove

 

SD card during acquisition. 

16.  During Acquisition, controls have been locked. To unlock them hold both the 

“Left”

 and 

“Right”

  

 

arrow keys simultaneously.

17.  When controls are unlocked signals can be viewed. Ending acquisition can be done on this  

 

 

screen by selecting the 

“Exit”

 tab. All result will be saved to SD card. 

WiFi Mode:

 

Wifi Mode allows the Mobile to stream acquisition data wirelessly to BioLab connected to a 

network. This stores data to computer rather than a SD card. To do this a BioNex must be connected 

to a computer with BioLab installed and open. See “BioLab Set-up” for more information.

 

 

1. Choose 

Menu

 from WiFi Mode main screen by selecting it with 

right context button “B”

.

2. 

This brings you to the 

Main Menu

 screen. Here you can change settings and power off the  

 

 

device by using the arrow keys and selection buttons.

04

Summary of Contents for MindWare Mobile

Page 1: ...Impedance Cardiograph WWW MINDWARETECH COM M a n u a l Product Manual Customer Support Email support mindwaretech com Phone 888 765 9735 ...

Page 2: ...ort Email support mindwaretech com Phone 614 626 4888 US Toll Free 888 765 9735 Fax 614 626 4915 Contents Product Overview p 01 Start up p 02 Local Mode p 02 04 Bio channel Scaling p 03 WiFi Mode p 04 06 Configuring BioLab p 06 Other Important Features p 06 Important Information Safety p 07 08 01 ...

Page 3: ...o move the outline around the Local or WiFi symbol and either clicking the Run or right context button B to finalize selection Local Mode Local mode stores data locally on any standard SD card inserted into the device The device includes an SD card but any other compatible SD card can also be used SDHC cards are recommended Data collected to SD card can be transferred to any computer with MindWare...

Page 4: ... and off by selecting the channel and using the Left and Right arrow keys to change between on and off 8 The gain of certain channels can also be edited by choosing the channel and using the Left and Right arrow keys on the Gain line found below On Off option See Bio channel scaling table below 9 It is important to note that on the right side of the Acquisition Settings screen an up or down arrow ...

Page 5: ... tab 17 At this point data is being recorded to the SD card Do not remove SD card during acquisition 16 During Acquisition controls have been locked To unlock them hold both the Left and Right arrow keys simultaneously 17 When controls are unlocked signals can be viewed Ending acquisition can be done on this screen by selecting the Exit tab All result will be saved to SD card WiFi Mode Wifi Mode a...

Page 6: ...k overview that the device is now using status will read Not Connected 9 Return to the main menu by selecting the Menu tab with the right context button B 10 From the Main Menu Screen select System Settings This screen contains Device Name Date Time as well as the About option 11 Selecting Device Name will bring up the option to edit the device name as it appears on the network Edit by using the a...

Page 7: ... device is now reading signals It is a good idea to preview the signals in BioLab before beginning data collection 21 During Acquisition controls have been locked To unlock them hold both the Left and Right arrow keys simultaneously 22 When controls are unlocked acquisition can be ended by selecting the End tab However it is recommended to control the acquisition from BioLab Other Important featur...

Page 8: ...SS standard s Operation is subject to the following two conditions 1 This device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Within the band 5150 5250 MHz the operation of this device is restricted to indoor use to reduce the potential for harmful interference to co channel mobile satellite system...

Page 9: ... or application designer Any use or application of MindWare Technologies LTD products for or involving medical or clinical treatment must be performed by properly trained and qualified medical personnel and all traditional medical safeguards equipment and procedures that are appropriate in the particular situation to prevent serious injury or death should always continue to be used when MindWare T...

Reviews: