3. 

On the 

Main Menu 

screen, choose either 

“Acquisition Settings”

 or 

“System settings”

 to view and  

 

edit settings. Going back to previous page can be done at any time by selecting the “

Back”

 tab   

 

with the 

left context button “A”

4. 

The device can also be shut off by selecting “Power Off” with the 

arrow keys

 on the 

Main Menu

   

 

screen and hitting the 

“Select”

 tab.

5. Choose 

“Acq Settings”

 to get to the Acquisition Settings screen. 

6. 

Acquisition #

 is assigned to distinguish data from each other. The device will auto-assign the    

 

last three digits to increment with each acquisition. The first four charcters are user-defined by   

 

selecting 

“Acquisition #”

 and using the 

arrow keys

 to adjust the number. When  

 

 

 

 

done, save changes with the 

“Save” 

tab.

7. 

Acquisition settings also lists the channels that the device can 

 

collect. These channels can be toggled 

on

 and 

off

 by selecting the 

 

channel and using the 

Left and Right arrow keys

 to change between

 

on

 and 

off. 

8. 

The gain of certain channels can also be edited by choosing the 

 

channel and using the 

Left and Right arrow keys

 on the 

“Gain”

 

 

line found below On/Off option. 

See Bio-channel scaling table 

below

9. 

It is important to note that on the right side of the Acquisition Settings screen an up or down    

 

arrow indicates that there are more selections not visible on screen but can be reached    

 

 

with the 

arrow keys

.

10.  Once settings for acquisition are complete, return to 

“Main Menu”

 and select “

System Settings”

11.  System Settings contains 

Date & Time Menu

 as well as the “

About”

 option. 

12.  On the 

Date & Time

 screen, device date and time that will be used in acquisition can be    

 

 

calibrated by using the arrow keys. Once date & time have been calibrated choose “

Save

”  

 

 

to record changes.

03

Gain

Norminal Range (V)

Theoretical Resolution (nV)

1X

+/- 2.4

286

2X

+/- 1.2

143

3X

+/- 0.8

95.4

4X

+/- 0.6

71.5

6X

+/- 0.4

47.7

8X

+/- 0.3

35.8

12X

+/- 0.2

23.8

Table 1: Bio-Channel Scaling

Summary of Contents for MindWare Mobile

Page 1: ...Impedance Cardiograph WWW MINDWARETECH COM M a n u a l Product Manual Customer Support Email support mindwaretech com Phone 888 765 9735 ...

Page 2: ...ort Email support mindwaretech com Phone 614 626 4888 US Toll Free 888 765 9735 Fax 614 626 4915 Contents Product Overview p 01 Start up p 02 Local Mode p 02 04 Bio channel Scaling p 03 WiFi Mode p 04 06 Configuring BioLab p 06 Other Important Features p 06 Important Information Safety p 07 08 01 ...

Page 3: ...o move the outline around the Local or WiFi symbol and either clicking the Run or right context button B to finalize selection Local Mode Local mode stores data locally on any standard SD card inserted into the device The device includes an SD card but any other compatible SD card can also be used SDHC cards are recommended Data collected to SD card can be transferred to any computer with MindWare...

Page 4: ... and off by selecting the channel and using the Left and Right arrow keys to change between on and off 8 The gain of certain channels can also be edited by choosing the channel and using the Left and Right arrow keys on the Gain line found below On Off option See Bio channel scaling table below 9 It is important to note that on the right side of the Acquisition Settings screen an up or down arrow ...

Page 5: ... tab 17 At this point data is being recorded to the SD card Do not remove SD card during acquisition 16 During Acquisition controls have been locked To unlock them hold both the Left and Right arrow keys simultaneously 17 When controls are unlocked signals can be viewed Ending acquisition can be done on this screen by selecting the Exit tab All result will be saved to SD card WiFi Mode Wifi Mode a...

Page 6: ...k overview that the device is now using status will read Not Connected 9 Return to the main menu by selecting the Menu tab with the right context button B 10 From the Main Menu Screen select System Settings This screen contains Device Name Date Time as well as the About option 11 Selecting Device Name will bring up the option to edit the device name as it appears on the network Edit by using the a...

Page 7: ... device is now reading signals It is a good idea to preview the signals in BioLab before beginning data collection 21 During Acquisition controls have been locked To unlock them hold both the Left and Right arrow keys simultaneously 22 When controls are unlocked acquisition can be ended by selecting the End tab However it is recommended to control the acquisition from BioLab Other Important featur...

Page 8: ...SS standard s Operation is subject to the following two conditions 1 This device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Within the band 5150 5250 MHz the operation of this device is restricted to indoor use to reduce the potential for harmful interference to co channel mobile satellite system...

Page 9: ... or application designer Any use or application of MindWare Technologies LTD products for or involving medical or clinical treatment must be performed by properly trained and qualified medical personnel and all traditional medical safeguards equipment and procedures that are appropriate in the particular situation to prevent serious injury or death should always continue to be used when MindWare T...

Reviews: