Batteries
The battery is charged whenever the patient monitor is
connected to an AC power source regardless of whether
or not the patient monitor is currently on. Whenever the
AC power is interrupted during patient monitoring, the
patient monitor will automatically run power from the
internal batteries.
For T6, T8 and T9, The patient monitor uses two battery
packs. If the two batteries have very different charge
capacity, the message [
Diff. Battery Voltages
] will be
displayed. In this case, apply AC power to the patient
monitor until the two batteries have approximately equal
charge capacity or are both fully charged. You cannot
use them before they have approximately equal charge
capacity or are fully charged.
Installing the Batteries
For T5, When the patient monitor uses two battery
packs, one battery pack can be easily exchanged
while the patient monitor operates from the other. If the
patient monitor uses one battery pack, you should insert
a new battery pack before the old one depletes.
For T5, battery compartment is located at the lower
right side. You can install the batteries by following this
procedure:
Turn off the patient monitor and disconnect the
power cord and other cables.
Place the patient monitor with its face up.
Open the battery compartment door.
Place the batteries into the slots.
Close the battery door and place the patient monitor
upright.
For T6, T8 and T9, battery compartments are located at
the bottom. The patient monitor uses two batteries. You
can install the batteries by following this procedure:
Turn off the patient monitor and disconnect the
power cord and other cables.
Place the patient monitor with its back on the desk.
Open the battery compartment door.
Place the batteries into the slots.
Close the battery door and place the patient monitor
upright.
1.
2.
3.
4.
5.
1.
2.
3.
4.
5.