9

Basic Operations

Turning Power On

Before  you  start  to  make  measurements,  check  the 
patient  monitor,  SMR  and  plug-in  modules  for  any 
mechanical  damage  and  make  sure  that  all  external 
cables, plug-ins and accessories are properly connected.

Plug  the  power  cord  into  the AC  power  source.  If  you 
run the patient monitor on battery power, ensure that the 

battery is sufficiently charged.

Press  the  power  on/off  switch  on  the  monitor’s  front. 
The  start-up  screens  are  displayed,  and  the  technical 
alarm  lamp  and  alarm  lamp  are  lit  in  blue  and  yellow 
respectively. Then,  the  alarm  lamp  turns  into  red,  and 
turns off together with the technical alarm lamp after the 
system gives a beep.

The monitor enters the main screen.

Starting Monitoring

Decide which measurements you want to make.

Connect  the  required  modules,  patient  cables  and 
sensors.

Check that the patient cables and sensors are correctly 
connected.

Check  that  the  patient  settings  such  as  [

Patient  Cat.

], 

[

Paced

], etc, are appropriate for your patient.

Refer to the appropriate measurement section for details 
of how to perform the measurements you require.

1.

2.

3.

4.

1.

2.

3.

4.

5.

Disconnecting from Power

To  disconnect  the  patient  monitor  form  the  AC  power 
source, follow this procedure:

Confirm that the patient monitoring is finished.

Disconnect the patient cables and sensors from the 
patient.

Make  sure  to  save  or  clear  the  patient  monitoring 
data as required.

Press and hold the power on/off switch for above 2 
seconds. The  patient  monitor  shuts  down  and  you 
can  unplug  the  power  cable.  If  the  patient  monitor 
does  not  shut  down  properly,  press  and  hold  the 
power on/off switch for 10 seconds to shut it down.

1.

2.

3.

4.

Note

:

Make  sure  that  the  operating  environment  of  the 

patient  monitor  meets  the  specific  requirements. 

Otherwise unexpected consequences, e.g. damage 
to the equipment, could result.

Do  not  use  the  patient  monitor  for  any  monitoring 
procedure on a patient if you suspect it is not working 
properly, or if it is mechanically damaged. Contact your 
service personnel or us.

Summary of Contents for BeneView T6

Page 1: ...BeneView T5 T6 T8 T9 Patient Monitor Quick Guide...

Page 2: ...Guide to help you with the basic operations For more details please refer to the Operator s Manual Note Before attempting to install operate or adjust this product please save and read the Quick Guide...

Page 3: ...patient monitor off An indicator is built in this switch It turns on when the patient monitor is on and turns off when the patient monitor is off 6 AC power LED It turns on when AC power is connected...

Page 4: ...uct Overview Side View T5 Bottom View T6 T8 T9 3 4 5 1 2 1 2 3 4 5 1 Compartment for CF storage card slot 2 Recorder 3 Battery compartment 4 Integral Module Racks 5 Integral Module Racks Side View T6...

Page 5: ...1 4 3 5 2 9 11 8 Radiator Product Overview Rear View T6 T8 T9 Rear View T5 1 2 3 4 5 6 8 9 10 11 2 7...

Page 6: ...ect the controlling devices USB mouse and USB keyboard of the secondary display 8 Digital Video Interface DVI It connects a standard digital video display 9 SMR Connector It connects the satellite mod...

Page 7: ...ards Vigilance II or VigileoTM monitor NMT module Neuromuscular transmission module SpO2 module Pulse oxygen saturation module BIS module Bispectral index module RM module Respiration mechanics module...

Page 8: ...Parameter Area A 7 Parameter Area B 8 Prompt Message Area 9 QuickKeys area Select screen items by pressing them direct ly on the patient monitor s screen You can enable or disable touchscreen operati...

Page 9: ...s and their waveforms will be displayed in area B Otherwise only the parameters will be displayed In Area C you can choose to display Timer and all the parameters whose associated waveforms will not b...

Page 10: ...Operations 1 2 3 4 Other menus are similar to the main menu and contain the following parts Heading gives a sum up for the current menu Main body displays options buttons prompt messages etc The menu...

Page 11: ...he patient settings such as Patient Cat Paced etc are appropriate for your patient Refer to the appropriate measurement section for details of how to perform the measurements you require 1 2 3 4 1 2 3...

Page 12: ...te setting for the screen brightness 10 is the brightest and 1 is the least bright If the patient monitor operates on battery power you can set a less bright screen to prolong the operating time of th...

Page 13: ...er adjustable Select the Volume Setup QuickKey or the ECG parameter window Others or the SpO2 parameter window Select QRS Volume or Beat Vol and then select the appropriate volume 0 means off and 10 i...

Page 14: ...rt Config menu select the configurations and User Maintenance Settings to export Then select the Export button A status message will report completion of the transfer 1 2 3 1 2 3 Basic Operations To i...

Page 15: ...use button as the primary button by following this procedure Select Main Menu Maintenance User Maintenance enter the required password Select Others to enter the Others menu Select Primary Button and...

Page 16: ...this procedure Select Main Menu Patient Data History Data Select a patient you want to view from the Patient Data List and then select Review Select Data Review 1 2 3 1 2 3 Note Data may be unable to...

Page 17: ...ill be triggered in the following conditions Timed recordings will start automatically at preset intervals If both Alarm and Alm Rec for a measurement are set on an alarm recording will be triggered a...

Page 18: ...pack can be easily exchanged while the patient monitor operates from the other If the patient monitor uses one battery pack you should insert a new battery pack before the old one depletes For T5 batt...

Page 19: ...charged indicates that the batteries are almost depleted and need to be charged immediately Indicates that no batteries are installed or only one battery is installed to the BeneView T8 or BeneView T9...

Page 20: ...Shenzhen Mindray Bio Medical Electronics Co Ltd Mindray Building Keji 12th Road South Hi tech Industrial Park Nanshan Shenzhen 518057 P R China Tel 86 755 81888998 Fax 86 755 26582680 www mindray com...

Reviews: