BeneLink Module Operator’s Manual
9-7
9.3.3
Setting the Serial Ports of the TcGas Monitoring Devices
Brand
Model
Protocol
Baud
Rate
Data
Bits
Stop
Bits
Parity
Remark
Radiometer
TCM
Tosca/CombiM
/TCM4/TCM40
MonLink
9600
8
1
Even
Setup(PW: 19100)
→
Techsettings
→
Continuous data output: MonLink
You need also the serial interface
adapter with cable from Radiometer.
(SN: 636-649)
SenTec
SenTec Digital
Monitor)
SenTecLink
115200
/
/
/
Interfaces
→
Serial Interface
→
Protocol: SenTecLink