4-68
BeneLink Module Operator’s Manual
4.15.2
Output Signals – Alarms
Monitor output: alarms from MAQUET SERVO-i/SERVO-s Ventilator
Priority
Label
Physiological alarms
High
Paw High
MV High
MV Low
Apnea
FiO
2
High
FiO2 Low
PEEP High
PEEP Low
Medium
EtCO2 High
EtCO2 Low
RR Low
RR High
Technical alarms
High
No Gas Supply Pressure
O2 cell disconnect
High
Technical
alarms
If the monitor shows the alarm [
High Technical
alarms
], the external device has one or more of the
following alarms:
Breathing system uP Module error
Inspiratory control uP Module error
Monitoring System uP Module error
Battery alarm
Power Failure
Mains Failure
O2 potentiometer error
CMV potentiometer error