4-74
BeneLink Module Operator’s Manual
Monitor output: alarms from MAQUET SERVO-U/SERVO-N Ventilator
Priority
Label
RR Low
Technical alarms
High
High
Technical
Alarm
If the monitor shows the alarm [
High Technical
alarms
], the external device has one or more of the
following alarms:
Battery alarm
Overrange alarm
Patient disconnected > 1 min
Time in waiting position > 2 min
Leakage too high
No Gas Supply Pressure
O2 cell disconnect
Patient Disconnected
Medium
Medium
Technical
Alarm
If the monitor shows the alarm [
Medium Technical
alarms
], the external device has one or more of the
following alarms:
CPAP high
CPAP low
Edi signal interference from ECG
Edi signal invalid
Expiratory cassette disconnected
Leakage High
No consistent patient effort
No patient effort
The nebulizer cannot be run on one battery
Battery in Use