MW170 Bench Meter
USB Export Selected
Logged data can be transferred separately by lots.
1. Press RCL to access the logged data.
LCD displays “PLEASE WAIT” followed by “LOG RECALL” with ACCEPT tag blinking and
the number of stored logs.
2. Press GLP/ACCEPT to confirm.
3. Use
keys to select the lot type (MANUAL, STABILITY or interval ###)
4. With the lot selected, press RANGE/
to export to USB flash drive.
LCD displays "PLEASE WAIT" followed by "EXPORTING" with ACCEPT tag and selected
lot name (MAN / STAB / ###) blinking.
LCD displays "EXPORTING" and the percentage counter, followed by "DONE" when
export is completed. Display returns to the lot selection screen.
Note: The USB flash drive can be safely removed if the USB icon is not displayed. Do not
remove the USB drive during export.
Overwriting existing data.
1. When the LCD displays "EXPORT" with ACCEPT and lot number blinking (USB icon
displayed), an identical named lot exists on the USB.
2. Press GLP/ACCEPT to continue. LCD displays "OVERWRITE" with ACCEPT tag blinking.
3. Press GLP/ACCEPT (again) to confirm. Not confirming exits the export.
Display returns to lot selection screen.
Data Management Warnings
“NO MANUAL / LOGS”
No manual records saved. Nothing to display.
“NO STABILITY / LOGS”
No stability records saved. Nothing to display.
"OVR" with lot ### (blinking) Identically named lots on USB. Select overwrite option.
"NO MEMSTICK"
USB drive is not detected. Data can not be transferred.
Insert or check the USB flash drive.
"BATTERY LOW" (blinking)
When low battery, export is not executed.
Recharge the battery.
Logged Data Warnings in CSV file
°
C ! Probe used beyond its operation specifications. Data not reliable.
°
C !! Meter in MTC mode.
°
C !!! Meter in NO TC mode. Temperature value only for reference.