EC / TDS
27
Note: Values above 1500 ppm (1500 mg/L) are only displayed in g/L unit.
The user can review the set values, date and time prior to measurement.
To scroll between information displayed on the third LCD line, use the
keys.
Note: If the reading is outside the range, the full-scale value is displayed blinking.
8.4. WARNINGS & MESSAGES
Messages Displayed During Calibration
y
If the reading exceeds expected value, “WRONG STANDARD” message is displayed
and calibration can not be confirmed. Check that correct calibration solution has
been used and / or clean the probe. See ELECTRODE CARE & MAINTENANCE section.
y
When using ATC mode, if the temperature of the solution is outside the accepted
interval, the “WRONG STANDARD TEMPERATURE” message is displayed. The °C tag
and the temperature are displayed blinking.