32
MW170 Bench Meter
9.4. WARNINGS & MESSAGES
Messages Displayed During Calibration
y
If an EC calibration is performed, the %NaCl calibration is automatically cleared.
A new %NaCl calibration is required.
y
If the reading exceeds the expected calibration standard, “WRONG STANDARD”
message is displayed and calibration is not confirmed.
Check if the correct calibration solution has been used and / or clean the probe.
See ELECTRODE CARE & MAINTENANCE section.
y
If the temperature is outside the 0.0 to 60.0 °C range, “WRONG STANDARD
TEMPERATURE” message is displayed. Temperature value is displayed blinking.
Messages Displayed During Measurement
y
If the salinity measurement exceeds the specified limits or the temperature exceeds
– 20 to 120°C (– 4.0 to 248.0 °F), “OUT OF SPEC” message is displayed.