pH
25
Repeat calibration after cleaning.
pH
mV
NO MTC
ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
Check Probe / Check Buffer
“CHECK PROBE CHECK BUFFER” is displayed when the electrode's slope exceeds the
highest accepted slope limit. Inspect the electrode and make sure the buffer solution is
fresh. Clean the probe to improve response time.
pH
mV
NO MTC
ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
pH
mV
NO MTC
ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
Bad Electrode
“BAD ELEC” is displayed if the cleaning procedure, performed after the above two
messages, is found to be unsuccessful. Replace the probe.
pH
mV
NO MTC
ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
Wrong Buffer Temperature
“WRONG TEMP” is displayed if the temperature of the buffer is out of range. The calibration
buffers are affected by temperature changes. During calibration, the instrument will
automatically calibrate to the pH value corresponding to the measured temperature but
compensate it to the value of 25 °C.