TROUBLESHOOTING
39
14. TROUBLESHOOTING
Symptom
Problem
Solution
Slow reponse /
Excessive drift
Dirty pH electrode
Soak the electrode tip in
MA9016
cleaning solution for 30 minutes
and then follow the Cleaning
procedure.
Reading fluctuates
up and down (noise)
Clogged/dirty junction.
Low electrolyte level
(refillable electrodes only)
Clean the electrode.
Refill with fresh
MA9012
electrolyte.
Display shows the
full scale value blinking
Reading out of range
Recalibrate the meter.
Check the sample is within
measurable range.
Check if electrolyte level and
general electrode status.
mV scale out of range
Dry membrane or dry
junction
Soak the electrode in
MA9015
storage solution for at least
30 minutes.
Display shows the
temperature unit blinking
Out of order temperature
probe
Replace the temperature probe.
Meter fails to calibrate or
gives faulty readings
Broken probe
Replace the probe.
LCD tags displayed
continuously at
startup
One of the keys is blocked
Check the keyboard.
If the error persists, contact
Milwaukee Technical Service.
“Er0, Er1, Er2” message
at start up
Internal error
Restart the meter.
If error persists, contact
Milwaukee Technical Service.