18
MW160
Bench Meter
7. SETUP
To configure the meter settings, modify default values or set measurement parameters:
y
Press and hold SETUP for about 3 seconds to enter Setup mode.
y
Use the
keys to navigate the menu (view parameters).
y
Press CAL to enter Edit mode (modify parameters). The selecied item is displayed
blinking.
y
Press RANGE to select between options.
For example, when setting current time, pressing RANGE switches between options
(hour, minutes) to be modified.
y
Use the
keys to modify values (value being modified is displayed blinking).
y
Press GLP/ACCEPT to confirm and save changes (ACCEPT tag is displayed blinking).
y
Press CAL to exit Edit mode without saving (return to menu).
y
Press SETUP to exit Setup mode.
The Setup menu items with options and default values are detailed here:
Item
Description
Options
Default
OFF
dAY
Calibration alarm time-out (pH range only)
When enabled, a warning will be displayed
after the set number of days from last
calibration has elapsed.
1 to 14 days or OFF
OFF
dISP
Display calibration buffers (pH range only)
When enabled, the buffers used in the
last calibration are displayed on the pH
measurement screen.
ON or OFF
ON
IonCG
Ion charge (ISE range only)
Use the following table to select the proper
ion charge.
–2, –1, 1, 2 or UndF
UndF
TIME
Current time
in hh:mm format
00:00 to 23:59
00:00