background image

5.2  

How to adjust

5.2.1  The transition to the adjustment mode

 

(1)  Press the power switch while holding down zero adjustment switch and the data hold switch at 

  

the same time.

 

 

(2)  Make sure that the LCD is lit all, press and hold the light 

  

switch over 1 second.

 

 

(3)  LCD will show the software version and CAL mark 

 

 

flashing, and then 2231-20 will move to the 

  

adjustment mode.

5.2.2    Each key operation in the adjustment mode

 

①  

Data hold switch: Change the adjustment point

 

②  

Zero adjustment switch: Run adjustment

5.2.3    DC current adjustment

 

(1)  Set to the adjustment mode (Measured value is shown 

 

 

in the main display, adjustment point is displayed in the 

  

sub-display).

 

(2) Connect the coil unit and current generator.

 

(3) Clamp the conductor of the coil unit.

 

(4) Ensure that the sub-display shows 0.00mA.

 

 

If the sub-display does not show 0.00mA, press the data hold switch, then the sub-display

 

 

will show 0.00mA.

 

(5)  Generate the current as same as the adjustment point.

 (6) 

After waiting for 10 seconds, press the zero adjustment switch, to perform the adjustment.

 

(7)  Press the Data Hold switch to move the next adjustment point.

 

(8)  Up to -100mA, repeat (5) (6) (7) procedure.              

 

  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  

Table 1 DC Current adjustment point 

adjustment 

procedure # 

adjustment point  

20mA range 

0mA 

+20mA 

0mA 

-20mA 

100mA range 

0mA 

+100mA 

0mA 

-100mA 

 

 

 

  

Summary of Contents for 2231-20 4-20

Page 1: ...ambient temperature of 23 C 2 C 73 4 F 3 6 F and relative humidity of 80 allow the fork meter to sit at this temperature for a minimum of 30 minutes before proceeding REPAIR SERVICE INSTRUCTIONS Calibration MILWAUKEE ELECTRIC TOOL CORPORATION 13135 WEST LISBON ROAD BROOKFIELD WISCONSIN 53005 2550 USA 58 92 2231d1 12 2012 2231 20 4 20 Milliamp Clamp Meter ...

Page 2: ... safety precautions to avoid injury and prevent damage to this product or products connected to it To avoid potential hazards use the product only as specified CAUTION These statements identify conditions or practices that could result in damage to the equipment or other property WARNING These statements identify conditions or practices that could result in personal injury or loss of life Specific...

Page 3: ...ry compartment cover and then remove the battery compartment cover 2 Remove batteries 14 4 2 Removal of Front panel 1 Loosen six screws 11 fixing the Front panel 2 Remove the battery harness 7 from PCB 4 3 Removal of PCB 1 Loosen one screw 6 fixing the PCB 2 Remove the PCB assembly 5 from Front panel 1 3 Remove the sensor module 16 from PCB 5 Adjustment Procedure Caution For your safety do not ope...

Page 4: ... Fig 1 Construction Drawing ...

Page 5: ...adjustment mode Measured value is shown in the main display adjustment point is displayed in the sub display 2 Connect the coil unit and current generator 3 Clamp the conductor of the coil unit 4 Ensure that the sub display shows 0 00mA If the sub display does not show 0 00mA press the data hold switch then the sub display will show 0 00mA 5 Generate the current as same as the adjustment point 6 A...

Page 6: ...or module Lack of the segment or abnormal lighting Defect of circuit components Replace the PCB LCD backlight or LED light is not lit Defect of circuit components Replace the PCB 7 List of Spare parts No Spare parts No Parts Name Qty Remark 18 9000018895 Lamp Button 2231 20 UL 1 17 9000019044 Lamp Button Bracket 2231 20 1 2 9000018894 Control Button 2231 20 UL 1 1 9000018890 Enclosure Front Case 2...

Reviews: