![Millipore ICW-3000 User Manual Download Page 44](http://html1.mh-extra.com/html/millipore/icw-3000/icw-3000_user-manual_1796917044.webp)
M
AINTENANCE
- 38 -
H
OW TO
R
ESET THE
UV
L
AMP
T
IMER
1.
Press the Main
and “+“ Buttons
together to enter
the menu. The
Display will
show C01.
2.
Press the Main
Button 3 times.
The Display will
show C03 and
“0” days left on
the UV Timer.
3.
Press the “+”
and “-“ Buttons
together. This
will reset the UV
Timer to 500
(days). The
Display will exit
the menu.
The UV Timer has been reset.